How do I set up an integration against Acteon Sopro PSPIX Generation 1 image plate scanner?
IAM Integration
PSPIX Generation 1 is a private-label Digora Optime machine. Please refer to our IAM Installation guide for image plate scanners:
How do I set up an integration against Soredex / Instrumentarium / Gendex / KaVo Panorama / OPG / Imaging Plate Scanner?
These devices often have very old firmware and will not work with modern IAM integration offered by DEXIS. IAM offers a optimized workflow and is a superior integration overall. Please contact your local dealer and let them assist you with the firmware upgrade.
TWAIN Configuration + Installation (Deprecated)
- Open the Muntra offline app, log in, and navigate to "This Computer" under the settings icon in the top right corner.
- Locate the "TWAIN PSPIX generation-1" device and open its settings. If you cannot see this device, please refer to the section Installing TWAIN Drivers below.
- Enable: Add to Menu
- Under Parameters, add the following:
- batch boolean TRUE
- flush boolean TRUE
- Reload the Muntra offline app.
Installing TWAIN Drivers
- Run the TWAIN PSPIX Setup.exe from the following link:
TWAIN PSPIX generation-1
- Proceed through the installation wizard without making any changes. Simply click "Next" and "Install."
- Complete the installation.
- Restart your computer.
- Open the Muntra offline app and navigate to the gear icon (top right) > This Computer.
- Verify that TWAIN PSPIX generation-1 appears in the device list, and then proceed to the previous section of this guide.